Instrument Details
- Make : Novocontrol Technologies Germany
- Model : Concept 80
- Installed on : March, 2009
- Purchased under the Scheme : ISPC Institute Funding.
Technical Specifications
- Temperature range : -150oC to 350oC
- RT to 800oC with Eurotherm controller for bulk samples with Alpha A analyzer only
- Frequency range :
- Novocontrol Alpha A Analyzer :- 3µHz to 20 MHz and
- Agilent E4991A RF analyzer :- 1MHz to 3GHz
- Measuring voltage : 0 to 3 Vrms / 70mA
- DC bias voltage : ±40V / 70 mA
- Impedance range : 0.01 Ohm to 100 TOhm
- Capacitance range : 0.001 pF to 1 F
- tan (δ) : 10-5 to 104
Working Principle
The Broadband dielectric spectrometer has two separate analyzers one below 20 MHz working on gain phase measurement method and other from 1 MHz to 3 GHz working on coaxial line reflectometry method.
- Permittivity, Modulus, Conductivity (AC), Specific resistance (resistivity), tan δ, Capacity, Impedance, Admittance, Inductivity, Measured temperature, measured time, Measured AC volt, Measured DC volt, measured AC current, Measured DC current.
- Magneto dielectric measurement (For Bulk and Thin Film sample)
- P-E Loop measurement (For Bulk and Thin Film sample)
- Piezoelectric Measurement (Only for Bulk sample)
The Broadband dielectric spectrometer has two separate analyzers one below 20 MHz working on gain phase measurement method and other from 1 MHz to 3 GHz working on coaxial line reflectometry method.
- Sample should be well polished and havingsmooth surface.
- Maximum of six samples will be accepted at a time for room temperature.
- Maximum of two samples will be accepted at a time with temperature measurement.
- Maximum sample dimension should be diameter of 3 cm and thickness 0.3 cm for Low frequency, and diameter 1 cm and 0.2 cm thickness for RF.
- After applying silver paste (Conducting coating) to samples it should be heated at 50oC at least for one hour.
- User has to bring liquid nitrogen.
- Only one registration will be accepted at a time.
- Click here to read instructions for Optimal Sample Geometry
Charges for external users apart from IITB's users
Organization Room temperature With temperature
Academic 200 1000
Industry 1500 5000